Determination of Picogram Levels of Diacerein in a Pharmaceutical Formulation by Flow-Injection Chemiluminescence
School of Pharmacy, Xi’an Medical University, Xi’an 710021, China
A simple, sensitive and inexpensive method for determination of diacerein by flow-injection chemiluminescence was proposed, based on the quenching effect of diacerein on the luminol-protein (bovine serum albumin, BSA) reaction. It was found that the decrement of CL intensity was linearly proportional to the logarithm of diacerein concentration ranging from 5.0 to 7.0×103 pg·mL-1 (r = 0.9968), with the limit of detection (LOD) of 1.0 pg·mL-1 (3σ). The proposed procedure was successfully applied to the determination of diacerein in pharmaceutical formulation, human saliva, and serum samples without interference from its potential impurities, with the recoveries ranging from 96.4% to 104.0% and the relative standard deviations (RSDs) less than 4.0% (n=6).
Key words: diacerein / luminol / chemiluminescence / folw injection
© The Authors, published by EDP Sciences, 2017
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.