Establishment of calibration equipment for defibrillator analyzer and evaluation of measurement uncertainty
1 Shanghai Institute of Measurement and Testing Technology, Shanghai, China
2 Shanghai Key Laboratory of On-line Testing and Control Technology, Shanghai, China
* Corresponding author: zhanzq@simt.com.cn
A cardiac defibrillator is a medical electrical device used to defibrillate the heart of patients. In JJF1149-2014, defibrillator analyzers are used to measure the output energy of cardiac defibrillators, whose energy accuracy is an important parameter. This manuscript describes establishment of calibration equipment for defibrillator analyzer, including voltage divider module, sampling module and data processing module. Measurement uncertainty is evaluated for measurement results at last.
© The Authors, published by EDP Sciences, 2021
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.